Ultrasound Imaging Overview

Ultrasound imaging uses echoes of high-frequency sound to see inside the body. It works like sonar: a probe sends a short pulse of sound into the body, then listens for the echoes that bounce back from tissue boundaries.

Learning Objectives

  • Describe sound as a longitudinal pressure wave, and define wavelength, frequency, amplitude, and speed
  • Explain how a piezoelectric transducer both sends ultrasound pulses and detects echoes
  • Explain how the timing of an echo sets its depth, and how the strength of an echo sets its brightness
  • Explain how acoustic impedance, reflection, refraction, and attenuation shape an ultrasound image
  • Explain the trade-off between resolution and depth when choosing a transducer frequency
  • Describe how the Doppler effect is used to image blood flow
  • Describe the main clinical applications, advantages, and limitations of ultrasound

Important Terminology

Term Definition
Ultrasound Sound with a frequency above the range of human hearing (above 20 kHz). Medical imaging uses about 2–15 MHz
Transducer The probe. It sends ultrasound pulses and converts the returning echoes into electrical signals
Piezoelectric crystal A material that vibrates when a voltage is applied, and produces a voltage when it is vibrated
Echo The part of a sound pulse that is reflected back to the transducer
Acoustic impedance (Z) How strongly a tissue resists the passage of sound: density × speed of sound
Attenuation The loss of sound energy as it travels through tissue
Echogenicity How many echoes a tissue sends back, which determines how bright it looks
Doppler effect The change in frequency of a wave reflected from a moving object

Useful Resources

Modality Comparison

Every imaging modality turns a different physical property into brightness:

Modality What sets the brightness of each pixel
Visible light The amount of light reaching the image sensor
X-ray and CT How strongly tissues attenuate the X-rays passing through them
MRI Proton density and the T1 and T2 relaxation properties of the tissue
Ultrasound The strength of the echoes reflected back from tissue boundaries

Sound Waves

Sound is a Pressure Wave

Sound is a mechanical wave: a wave of pressure that travels through a medium such as air, water, or tissue. A vibrating source (a speaker, a tuning fork, or a crystal in an ultrasound probe) pushes on the molecules next to it, and they push on their neighbors.

longitudinal wave

A sound wave. Molecules bunch together in zones of high pressure (compression) and spread apart in zones of low pressure (rarefaction). The pressure curve below shows the same wave as a graph

longitudinal wave animation

A longitudinal wave in motion. The wave travels from left to right, but each molecule only moves back and forth around its starting point. Keep your eye on one red dot (Animation: Dan Russell, Penn State)

Key ideas

  • The wave travels, but the molecules only move back and forth
  • Sound needs a medium. It cannot travel through a vacuum

Transverse vs Longitudinal Waves

transverse vs longitudinal wave

Longitudinal and transverse waves. In a longitudinal wave, like sound, the medium moves back and forth along the direction of travel, forming compressions and expansions (rarefactions). In a transverse wave, like light, the wave oscillates at right angles to the direction of travel

  • Electromagnetic waves (light, X-rays, radio waves) are transverse waves
  • Sound waves in air and soft tissue are longitudinal waves

Properties of Waves

wave properties

Properties of a wave

Property Definition
Wavelength (λ) The distance between two neighboring peaks (or two compressions)
Frequency (f) The number of complete cycles that pass a fixed point each second, measured in hertz (Hz). We hear frequency as pitch
Amplitude The height of the pressure peaks. We hear amplitude as loudness
Phase Where in its cycle the wave is at a given point in time
Speed (c) How fast the wave travels through the medium

wave phase

Phase. Two waves with the same frequency and amplitude can be shifted relative to each other: they are out of phase

Speed, frequency, and wavelength are linked by a simple equation:

\(c = f \lambda\)

Where:

  • \(c\) is the speed of sound, in meters per second (m/s)
  • \(f\) is the frequency, in hertz (Hz)
  • \(\lambda\) is the wavelength, in meters (m)

The medium sets the speed

The speed of sound depends on the medium, not on the frequency: about 343 m/s in air and about 1,540 m/s in soft tissue. So in a given tissue, raising the frequency shortens the wavelength. At 5 MHz in soft tissue, the wavelength is about 0.3 mm.

Ultrasound

Frequencies Above Our Hearing

Humans hear sounds from about 20 Hz to 20 kHz (20,000 Hz), and the upper limit drops with age. Sound below 20 Hz is called infrasound. Sound above 20 kHz is called ultrasound.

sound spectrum

The sound spectrum. Elephants communicate with infrasound, humans hear from 20 Hz to 20,000 Hz, and bats use ultrasound. Medical imaging uses frequencies in the megahertz (MHz) range, far above our hearing (Image: howequipmentworks.com)

Many animals use ultrasound. Bats find prey by echolocation, with calls of about 20–120 kHz, and dolphins echolocate with clicks up to about 150 kHz. Medical ultrasound uses the same idea: send out a sound and listen for the echo.

Advantages and Disadvantages

Diagnostic imaging uses frequencies of about 2–15 MHz.

Advantages Disadvantages
Painless and non-invasive Blocked by air (lungs, bowel gas) and by bone
No ionizing radiation Higher frequencies cannot image as deep
Real-time imaging Lower resolution than CT or MRI at depth
Inexpensive, portable, and widely available Small field of view
The beam can be steered and focused Image quality depends on the operator's skill and on the patient

The Transducer

The transducer (or probe) is both the speaker and the microphone. It sends short pulses of ultrasound into the body, then listens for the echoes. It spends far more time listening than sending.

Piezoelectric Crystals

The key component of a transducer is an array of piezoelectric crystals. These materials have a special property:

  • When a voltage is applied, the crystal changes shape. An alternating voltage makes it vibrate and send out sound
  • When an echo squeezes the crystal, it produces a small voltage, which becomes the signal

piezoelectric crystal

The piezoelectric effect works in both directions. Left: pressure or vibration on the crystal produces an electrical current (receiving an echo). Right: an applied voltage changes the crystal's shape (sending a pulse) (Source)

The same crystal both sends and receives, so the cycle runs like this: send a pulse, listen for echoes, repeat.

Piezoelectricity in everyday life

A push-button gas lighter uses the same effect: pressing the button strikes a piezoelectric crystal, which produces a high-voltage spark.

Types of Transducers

Different jobs need different transducers, with different frequencies and footprint shapes.

Probe Typical frequency Image shape Used for
Linear High, ~5–15 MHz Rectangle Shallow structures: vessels, thyroid, muscles, needle guidance
Curvilinear (convex) Low, ~2–5 MHz Wide fan Deep structures: abdomen, obstetrics, trauma (FAST) exams
Phased array Low, ~1–5 MHz Narrow fan from a small point The heart: the small footprint fits between the ribs
Endocavitary ~5–10 MHz Varies Imaging from inside the body: transesophageal, endoscopic, transvaginal

transducer imaging field

The imaging field. Each image is a thin slice through the body. Its shape depends on the probe: a fan (like a slice of pizza) for curvilinear and phased-array probes, or a rectangle for linear probes

Making an Image from Echoes

Pulse-Echo Imaging

The machine sends a pulse and starts a clock. Every echo that comes back tells it two things:

The two rules of pulse-echo imaging

  1. Echo time → depth. The longer an echo takes to return, the deeper its source
  2. Echo strength → brightness. Stronger echoes are drawn as brighter pixels

To turn time into depth, the machine assumes that sound always travels through soft tissue at 1,540 m/s (1.54 mm per microsecond). The pulse makes a round trip, down and back, so:

\(\text{depth} = \frac{c \times t}{2}\)

Where:

  • \(c\) is the speed of sound in soft tissue (1.54 mm/µs)
  • \(t\) is the time between sending the pulse and receiving the echo

Example: an echo that returns after 130 µs came from 1.54 × 130 ÷ 2 ≈ 100 mm (10 cm) deep.

Only a tiny fraction of the sound comes back: about 1/10,000 of the transmitted power returns to the transducer.

A-mode and B-mode

pulse echo

From echoes to an image. Top: an A-mode trace plots echo amplitude against time. Each echo's arrival time gives the depth of a tissue boundary. Bottom: in B-mode, the same echoes are drawn as bright dots at their depths. Deeper echoes tend to be weaker, because the sound loses energy on the way down and back

  • A-mode (Amplitude mode) is a one-dimensional plot of echo strength against time (or depth). It measures distances but does not form an image, and today it is used mainly to measure the length of the eye.

A-scan

An A-scan of the eye. The probe sends pulses along a single line through the eye. Each spike in the trace is an echo from a boundary, such as the lens or the retina, and its position along the trace shows how deep that boundary is

  • B-mode (Brightness mode) turns each echo into a dot whose brightness matches the echo's strength. The transducer sweeps the beam across many lines side by side, building up a two-dimensional, cross-sectional image. This is the standard grayscale ultrasound image.

B-scan

Building a B-scan. Each trace (scan line) records a row of dots at the depths of its echoes. Many traces fanned out side by side outline the structures in a cross-sectional image

Describing Brightness

Radiologists and sonographers use a special vocabulary to describe how bright a structure looks:

Term Appearance Examples
Anechoic Black: no echoes Simple fluid: bladder, gallbladder, cysts, blood in vessels
Hypoechoic Darker than the surrounding tissue Many masses and lymph nodes
Isoechoic The same brightness as the surrounding tissue
Hyperechoic Brighter than the surrounding tissue Bone surfaces, stones, fat, tendons

How Tissues Interact with Sound

How a tissue interacts with sound determines what the image looks like: where echoes come from, how bright they are, and what gets lost along the way.

Acoustic Impedance

Acoustic impedance (Z) is a physical property of a tissue that describes how strongly it resists the passage of sound:

\(Z = \rho \times c\)

Where \(\rho\) is the density of the tissue and \(c\) is the speed of sound in it. Impedance is measured in rayls (the values below are in megarayls, MRayl).

Echoes form wherever impedance changes, at the boundary between two tissues. The bigger the change, the stronger the echo. The fraction of the sound reflected at a boundary is:

\(R = \left(\frac{Z_2 - Z_1}{Z_2 + Z_1}\right)^2\)

Tissue Z (MRayl)
Air 0.0004
Lung ~0.18
Fat ~1.38
Blood, liver, muscle ~1.6–1.7
Bone ~7.8
Boundary Sound reflected What you see
Liver / muscle ~0.03% A subtle boundary
Fat / muscle ~1% A clear boundary; most sound passes through
Soft tissue / bone ~43% A bright surface with a shadow behind it
Soft tissue / air ~99.9% A bright line, and nothing beyond it

Why we use gel

Almost all sound is reflected at a tissue–air boundary. Even a thin layer of air between the probe and the skin would block the sound, so we spread coupling gel on the skin to push the air out. For the same reason, ultrasound cannot see through the lungs or gas in the bowel.

strong reflectors

Strong reflectors cast acoustic shadows. The bile inside the gallbladder is anechoic (black). The gallstones are bright, and because they block the sound, they leave a dark acoustic shadow below them. Notice also the layers of the abdominal wall at the top and the grainy texture of the liver (Image: Mikael Häggström, CC0)

Reflection and Scattering

Sound has to come back to the transducer to be seen. Tissues return sound in two ways:

  • Specular reflection happens at large, smooth boundaries, such as the surface of an organ or a bone. These boundaries act like mirrors: sound leaves at the same angle it arrived, like a bank shot in pool. A boundary that faces the probe sends a strong echo straight back. A boundary tilted at an angle sends the sound away, so the echo never returns and that region looks dark
  • Scattering happens at structures smaller than the wavelength, such as cells and small blood vessels inside an organ. They send weak echoes in all directions, which give organs their grainy, speckled texture

Strong reflectors also leave little sound to continue deeper, so the tissue behind them is dark. This is called acoustic shadowing.

Why is blood dark?

Blood looks black (anechoic) not because its impedance is low (it is close to that of soft tissue) but because it is uniform and contains few structures large enough to scatter sound at these frequencies.

Disease can change how a tissue scatters sound. For example, a fatty liver is brighter than a normal liver, and the sound fades faster with depth.

Partial Reflection and Refraction

At every boundary, part of the sound is reflected back toward the probe and part is transmitted deeper. In practice, there are many boundaries, so each pulse produces a series of echoes.

partial reflection

Partial reflection. At a boundary, part of the wave is transmitted and part is reflected. The reflected wave (traveling to the left) is the echo detected by the transducer

Refraction is the bending of a wave when it crosses, at an angle, into a medium where it travels at a different speed. Sound, like light, can refract. The bigger the change in speed, the more it bends. A wave that hits a boundary straight on (at 90˚) does not bend.

Refraction matters because the machine assumes every echo traveled in a straight line. Refraction can therefore put structures in the wrong place, duplicate them, or cast thin dark edge shadows at the sides of round, fluid-filled structures.

Attenuation

Attenuation is the loss of signal strength as sound travels through tissue. It has three causes:

  1. Absorption: sound energy is converted to heat. This is the largest contributor
  2. Reflection: energy sent back at boundaries does not continue deeper
  3. Scattering: energy sent off in other directions

Attenuation is measured in decibels (dB). A useful rule of thumb is that soft tissue attenuates about 0.5 dB per cm per MHz (one way). Because attenuation grows with frequency, a high-frequency beam fades much sooner than a low-frequency beam.

Tissue Attenuation (dB/cm at 1 MHz)
Water 0.002
Blood 0.18
Fat 0.63
Liver 0.5–0.94
Kidney 1.0
Muscle 1.3–3.3
Bone 5.0 or more

Fluid barely attenuates sound, so tissue behind a fluid-filled structure (like the bladder or a cyst) receives more sound than its neighbors and looks brighter. This is called posterior acoustic enhancement, and it helps identify cysts.

Decibels

Decibels are a relative, logarithmic measure. Every 10 dB is a factor of 10 in intensity:

Change Intensity ratio
3 dB About 2×
10 dB 10×
20 dB 100×
60 dB 1,000,000×

A logarithmic scale is useful because ultrasound echoes span an enormous range of strengths, from strong bone echoes to the faint scatter inside an organ.

Compensating for Attenuation: Gain and TGC

Echoes from deep tissue are weaker simply because they traveled farther. The machine compensates with gain (amplification):

  • Overall gain amplifies all echoes equally, making the whole image brighter or darker
  • Time gain compensation (TGC) adds more amplification to echoes that arrive later, because they came from deeper and lost more energy. TGC is adjusted with a set of sliders on the console, one for each depth, so that deep tissue isn't too dark

Choosing a Frequency: Resolution vs Depth

Choosing a transducer frequency is a trade-off:

  • Higher frequency → shorter wavelength → better resolution, but more attenuation, so it cannot image as deep
  • Lower frequency → deeper penetration, but lower resolution

attenuation and depth

Higher frequencies lose signal faster. Round-trip loss at 0.5 dB/cm/MHz. If the machine can detect echoes that are up to 60 dB weaker than the transmitted pulse, a 3 MHz probe can image about 20 cm deep, but a 15 MHz probe only about 4 cm

So we choose the highest frequency that still reaches the depth we need: high frequencies for the thyroid, tendons, or vessels near the skin, and low frequencies for the abdomen or the heart.

Doppler Ultrasound

The Doppler effect is the change in frequency of a wave when its source or reflector is moving. When a source moves toward you, the waves bunch up and you hear a higher pitch. When it moves away, the waves spread out and you hear a lower pitch. Think of the siren of a passing ambulance.

Ultrasound uses the Doppler effect to detect motion, most often blood flow. Echoes from blood moving toward the probe return at a slightly higher frequency, and echoes from blood moving away return at a slightly lower frequency.

Color Doppler paints this frequency shift in color, inside a box (the color window) on top of the grayscale image:

  • By convention, red = toward the transducer and blue = away. Remember it as BART: Blue Away, Red Toward
  • Always check the color bar on the screen, because the color map can be inverted

color Doppler of the carotid artery

A carotid artery Doppler exam. On the screen, color Doppler (right) paints the blood flowing through the internal carotid artery, and spectral Doppler (left) plots blood velocity over time, with a peak at each heartbeat. On the console below, the column of small sliders is the TGC control (Image: Etan J. Tal, CC BY 3.0)

Common misconceptions

  • Color does not mean artery or vein. It only shows the direction of flow relative to the probe
  • Doppler only detects motion along the beam. Flow at 90˚ to the beam shows little or no color, so the angle of the probe matters

Reading Ultrasound Images

You really need to know your anatomy to read ultrasound. Unlike CT and MRI, ultrasound images are not limited to axial, sagittal, and coronal planes: the probe can be angled in any direction. This is a strength, because you can line up with any structure, and a challenge, because you have to work out the orientation of every image. Each clinical exam has its own set of standard views (for example, the parasternal long-axis view of the heart).

Ultrasound is a tomographic modality, like CT and MRI: each image is a slice through the body, rather than a projection like an X-ray. Unlike CT and MRI, ultrasound slices are acquired in real time, by hand.

Clinical Applications

Obstetrics

Because it uses no ionizing radiation, ultrasound is the imaging method of choice during pregnancy. It is used to date the pregnancy, check fetal growth and anatomy, and locate the placenta.

fetal ultrasound

Obstetric ultrasound showing the profile of a 20-week fetus (Image: Goleisureintl, CC BY 4.0)

Emergency Medicine (FAST)

The Focused Assessment with Sonography in Trauma (FAST) exam is a quick bedside scan that checks for free fluid (usually blood) around the heart and in the abdomen of an injured patient. Free fluid is anechoic, so it appears as a black stripe between organs.

Guided Procedures

Ultrasound lets a clinician watch a needle in real time, making injections, biopsies, and catheter placement safer and more accurate.

Endoscopic Ultrasound

Endoscopic ultrasound (EUS) puts a small, high-frequency probe on the tip of an endoscope. From inside the stomach or duodenum, it can image nearby structures, such as the pancreas and bile duct, without bowel gas in the way. A thin needle passed through the endoscope can sample a lymph node or mass under direct ultrasound view (fine needle aspiration).

Echocardiography

Echocardiography images the beating heart in real time, along with blood flow through its chambers and valves.

  • Transthoracic echo (TTE) uses a probe on the chest, imaging through the gaps between the ribs
  • Transesophageal echo (TEE) passes a probe down the esophagus, right behind the heart, for a clearer view of structures such as the valves and left atrium

parasternal long axis echocardiogram

The parasternal long-axis view, a standard echocardiography view. The probe sits on the chest beside the sternum, and its imaging plane (blue) slices lengthwise through the left ventricle, the mitral valve, and the aortic valve (Illustration: Patrick J. Lynch, via CardioNetworks ECHOpedia, CC BY-SA 3.0)

3D Ultrasound

A conventional probe sweeps a flat, 2D fan. A 3D (matrix-array) probe steers the beam in two directions to fill a pyramid-shaped volume, which can be rendered in 3D, for example to view a fetus or the heart valves.

Summary

Strengths Limitations
Non-invasive and non-ionizing Cannot see through air or bone
Real-time and portable Trade-off between resolution and depth
Shows anatomy and function, such as blood flow and heart motion Depends on the operator's skill